Introduction
Methamidophos is a tioorganophosphate used as an acaricide and insecticide. It inhibits cholinesterase and is employed in agriculture since 1970 [6]. Methamidophos is used for insect control in alfalfa, potato, cotton and tomato crops.
Methamidophos is highly toxic when ingested orally, through the skin or by way of the respiratory tract [5] [7] and the information related to the distribution and metabolization of methamidophos in human beings is scarce. It is eliminated mainly by way of the urinary system either without transformation or as its metabolite O,S-Dimethylhydrogen phosphorothioate (O,S-DMPT) [6] [8]. Its main function is to inhibit the acetylcholinesterase enzyme thus preventing the degradation of the acetylcholine which produces a cholinergic hyperactivity which in turn puts the individual in a state of lethal risk.
Methamidophos has been associated with acute fatal intoxications in Colombia [4]. This makes its determination in biological matrices a very important issue in the field of forensic toxicology in the country.
For this study a quick and simple method was developed in order to quantify the methamidophos in the blood and its qualitative identification in urine using LC-ESI-MS/MS. This analytical methodology was applied to a case of fatal intoxication suspected of the presence of pesticides.
Materials and methods
Methamidophos and methamidophos-d6 standards were acquired from Dr Ehrenstorfer GmbH. All the reagents were analytical grade except methanol which was HPLC grade. The methamidophos and methamidophos-d6 stock solutions were prepared at a concentration of 1 mg/mL in methanol. The blanks for urine were acquired from volunteers. The blank for the blood was prepared from a 50:50 concentrated red cell solution and deionized water.
Procedure
Blood: One milliliter of the whole blood sample was pipetted into a test tube and 50 µL of methamidophos-d6 (Internal standard) were added at a concentration of 5,0 µg/mL and then stirred. 1 mL of cold acetonitrile was slowly added drop by drop (at freezer temperature) while in the vortex (It is essential not to stop the stirring process). The test tube was sealed and left in the freezer during 30 minutes. Then it was centrifugated at 2000 rpm and the organic fraction was transferred to a new test tube. The solution was evaporated- while being stirred- at 45ºC to a final volume of 0.5 mL. It was then evaporated under vacuum until fully dry.
Finally the analyte was reconstituted with 100 µL of a methanol/water solution 50:50 (v/v) with 0,1% formic acid. This procedure was applied to each level of the calibration curve.
Urine: 100 µg L of methamidophos-d6 solution was added to a 2 mL-urine sample, it was then stirred and extracted by a solid phase procedure. In this procedure a Hypersep verify CX 200mg/6mL cartridge was prepared by adding 2,0 mL of methanol first followed by 2,0 mL of a pH 6.0 phosphate buffer. Once the urine sample was added to the cartridge it was washed with 2,0 mL deionized water, 0,5 mL acetic acid 0,01M, and 100 µL methanol then it was dried under vacuum during 10 minutes. The dried sample was then eluted with the following series of solvents: 1,5 mL acetone-chloroform 50:50 (v/v), mL acetone /dichloromethane 50:50 (v/v), 1,5 mL ethyl acetate/ammonium hydroxyde 98:2 (v/v) and finally 1,5 mL dichloromethane/isopropanol/ammonium hydroxyde 78:20:2 (v/v/v). The eluate was dried at 45ºC and reconstituted with 100 µL of a methanol/water solution 50:50 (v/v) with 0,1% formic acid.
Liquid chromatography-mass spectrometry conditions
The LC-MS system used consisted of a Thermo Electron Corporation model Thermo Surveyor. The chromatographic separation was performed on a HPLC Hypersil Gold PFP column (50 mm x 2,1 mm; 5 µm). The column temperature was 40ºC. The mobile phase was a gradient of a mixture of methanol with 0.1% formic acid and a solution of ammonium formate at a constant flow (200 µL/min) programmed as follows: Initial 2% methanol for 1 min., then 2-98% methanol for 9-13 min. and finally, decreased to 98-2% methanol for 14-16 min.
All the mobile phase solvents were previously vacuum filtered using a membrane with a pore size of 0,22 µm. The injection volume was 10 µl.
MS Detection
Detection was performed on a LCQ Advantage Max ion trap mass spectrometry in tandem in product ion scan mode. The detector was equipped with an electrospray ionization source (ESI) in positive mode. Conditions were optimized for methamidophos and methamidophos-d6 by injecting them apart using a syringe pump in “T” (table 1).
This system permits the mixing of the pesticide previously prepared at a concentration of 2,0 µg/mL with the mobile phase at a 50:50 proportion. The optimized conditions of the MS were as follows:
capillar voltage 9,00 V, source voltage 5,00 kV, capillar temperature 160ºC, lens voltage 5 V, sheath gas flow rate 55 units and auxiliary gas flow rate 15 units.
| Compound | Transition (m/z) | CE (%) | Polarity | RT |
|---|---|---|---|---|
| Methamidophos | 141,9→111,9 141,9→93,9 | 30 | positive | 2,72 |
| Methamidophos-d6 (Internal Standar) | 148,1→114,9 148,1→96,9 | 30 | Positive | 2,59 |
Tab. 1 shows the values for the ions (m/z), collision energy (CE), polarity and retention time, all of them optimized for the identification of each compound.
Case history
A 17 years old boy was found dead in his home.
The authorities suspected an exogenous intoxication with a pesticide. Samples of urine, blood and gastric contents were taken during the autopsy. The gastric content was analyzed by GC/MS yielding positive results for methamidophos.
Results
Figure 1 shows the chromatogram and Figures 2 and 3 the mass spectra for methamidophos and its internal standard in the blood sample. The LOD determined by successive dilution of the compound in the matrix was 50 ng/mL for methamidophos and methamidophos-d6 in blood and urine.



The lineal range for the calibration curve was between 0,10 and 5,00 µg/mL with a correlation coefficient of 0,9901. Table 2 shows the data for precision (CV) -evaluated as repeatability and intermediate precision (same day, two analysts and same analyst, two days)- accuracy (relative main error, RME) and recovery percentages evaluated at levels 0,10, 0,25, 2,50 and 5,00 µg/mL
| Analite | Concentratio n (µg/mL) | Repetibility (n=5) | Accuracy (n=5) | Intermediate precision Analyst (n=10) | Accuracy (n=10) | Intermediate precisión Analysis day (n=10) | Accuracy (n=10) | Recovery percentages |
|---|---|---|---|---|---|---|---|---|
| C.V. | M.R.E. | C.V. | M.R.E. | C.V. | M.R.E. | % | ||
| Methamidophos | LOQ = 0,10 | 12,17 | -9,9E-03 | 9,84 | -2,5E-03 | 13,20 | 3,6E-03 | 97,1 |
| 0,25 | 7,43 | 4,4E-03 | 5,63 | -0,2E-03 | 6,72 | -25,3E-03 | 90,6 | |
| 2,50 | 1,91 | 0,3E-03 | 2,02 | 1,1E-03 | 1,63 | -7,9E-03 | 98,4 | |
| ULOQ = 5,00 | 4,61 | -49,4E-03 | 4,04 | -25,1E-03 | 3,46 | -34,6E-03 | 92,9 |
Tab. 2. Data for precision (CV), accuracy (relative main error, RME) and recovery percentages.
The matrix effect evaluated by the blood post-extraction addition experiment, n=6 at 0,25 µg/ml, was 14,75 % with a CV of 9,6 while that for a concentration of 5,0 µg/mL was -14,15% with a CV of 13,9.
The above methodology was applied to a real forensic case in which the methamidophos concentration in blood for the victim was found to be above the upper quantification limit. It was reported as being higher than 5,0 µg/mL.
Methamidophos was found in the urine.
Discussion and conclusions
This study established a method for the determination of methamidophos in blood and urine by LC-ESI-MS/MS. The method’s recovery percentages were higher than 90% and repeatability, intermediate precision, accuracy, matrix effect and linearity values comply with the acceptation criteria established by bioanalytical guidelines [1]
LC-ESI-MS/MS is a sensible technique for the detection of methamidophos in biological samples as compared with results by GC/MS previously reported [2] [3]. The method shows high selectivity and allows for the unequivocal determination of methamidophos in biological samples, a fundamental condition for the forensic toxicology laboratories.
Even though methamidophos was found in blood at a concentration higher than 5.0 µg/mL quantifying it accurately was not possible due to the small sample size available, nevertheless the analysis of the urine sample confirmed the presence of the pesticide in the victim. Both results support the hypothesis that the cause of the death was poisoning with methamidophos.
Acknowledgments
We would like to express our grateful appreciation to Departamento de Toxicología Universidad Nacional de Colombia specially to Dr Jairo Tellez, also to Laboratorio de Toxicología, Instituto Nacional de Medicina Legal y Ciencias Forenses, specially to Consuelo Mora and finally to Mr. José Peláez for the translation.
References
- EMEA/CHMP/EWP. (2011). Guideline on bioanalytical method validation. European Medicines Agency, (July 2011). ↑
- Adachi, N., Kinoshita, H., Nishiguchi, M., Takahashi, M., Ouchi, H., Minami, T., Matsui, K., et al. (2008). Simultaneous analysis of acephate and methamidophos in human serum by improved extraction and GC-MS. Forensic Toxicology, 26(2), 76–79. ↑
- Adachi, N., Kinoshita, H., Nishiguchi, M., Takahashi, M., Ouchi, H., Minami, T., Matsui, K., et al. (2011). Determination of acephate and methamidophos in tissues: appearance of matrix effect in gas chromatography–mass spectrometry. Forensic Toxicology, 29(2), 159–162. ↑
- Castro, H., & Mora, M. (2006). Validación Metodología para análisis de plaguicidas por Gases-Masas en muestras forenses. Rev. Instituto Nacional de Medicina Legal y Ciencias Forenses, 19(3), 24–33. ↑
- Krieger, R. (2010). Hayes’ Handbook of Pesticide Toxicology (Third Edit., p. Cap 12, 73,75). ↑
- Moffat, A., & Osselton, D. (2011). Clarke’s Analysis of Drugs and Poisons: in pharmaceuticals, body fluids, and postmortem material (pp. 258–287, 843, 1651.) ↑
- Olson, K. R. (2004). Poisoning & Drug Overdose (Fourth edi., p. Section II y IV). ↑
- Tomaszewska, E., & Hebert, V. R. (2003). Analysis of O,S-dimethyl hydrogen phosphorothioate in urine, a specific biomarker for methamidophos. Journal of agricultural and food chemistry, 51(21), 6103–9. ↑